Overview
The HK-0103 Polymer-Containing Oil Shear Stability Tester (Diesel Nozzle Method) meets the requirements of test methods SH/T 0103 and ASTM D6278. It is suitable for determining the shear stability of polymer-containing oils using the diesel nozzle method, specifically measuring the kinematic viscosity loss of polymer-containing oils at 100℃ with a European diesel nozzle tester. The viscosity loss reflects the degradation degree of the polymer under the shearing action of the nozzle.
Functional Features
- The accuracy of testing standard samples with this instrument is ≥98%. For the detection method, conduct a test using the standard oil provided by the manufacturer, and the viscosity drop value shall be between 2.75 and 2.85.
- The difference between two test results obtained by the same operator, using the same instrument and testing the same sample under the same operating conditions, shall not exceed 1.05%.
- The nozzle used in the instrument is an imported Bosch DN8S2 type needle bolt nozzle from Germany.
- Diesel dual-plunger injection pump: imported Bosch PE2A90 from Germany.
- Equipped with a time relay to control the working time of the diesel dual-plunger injection pump.
- Users can optionally equip a low-temperature circulating water cooling system according to their needs, with the temperature range of ±5℃.
Technical Parameters
- Control Power Supply: AC 380V 50Hz 3kW
- Motor Power Supply: AC 380V 50Hz 1.1kW
- Pressure Gauge: Imported anti-vibration type, 0~40MPa
- Thermometer: Display accuracy ±0.1℃