Overview
HKT-010 smoke point heat pipe calibration device, in line with the requirements of JJG 130 calibration regulations, applicable to a variety of smoke point temperature
sensor and temperature instrumentation calibration, can replace the use of oil tank, salt tank and other thermostatic tank. It can be used instead of other constant
temperature baths such as oil bath and salt bath. It can be used for the calibration of fully-immersed and locally-immersed bimetal thermometers.
Type 1 Measuring range: 60℃~120℃; Type 2 Measuring range: 100℃~200℃.
Functional characteristics
1. Small temperature fluctuation, uniform and stable temperature field, with its own pressure stabilization system
2. Fast heating speed, from room temperature to 200 ℃ only 40min
3. No need to replace the working medium, the system is fully sealed, no noise, no fumes, no pollution of the environment, no harm to the body of the inspector, is a green smoke point detection devices
4. Convenient to operate, once the temperature is set, the automatic operation of temperature control
5. Long service life, generally can be used for more than 10 years
Technical parameters
1. Temperature fluctuation and temperature field uniformity
Temperature fluctuation: 60 ℃ ~ 120 ℃; ≤ 0.04 ℃
Temperature field uniformity:
a. The working area of the horizontal temperature difference ≤ 0.02 ℃
b. Maximum temperature difference in the working area ≤0.04℃.
Temperature fluctuation: 100℃~200℃;≤0.04℃
Temperature field uniformity:
a. Work area horizontal temperature difference ≤ 0.02 ℃
b. Maximum temperature difference in the working area ≤ 0.04 ℃
2. Temperature stabilization time; ≤40min
3. Effective diameter and length of temperature measuring tube: 20mm; 360mm
4. Number of temperature measuring tubes: 4
5. Temperature control sensor: Pt100 platinum resistance



Thermometer Name Temperature Measurement Range